3
Selecting the Right Flow Sensor
The following pictures are guidelines for correct installation of air velocity trans-
mitters.
Reliable measure-
ments can be made
by positioning the
transmitter after fil-
ters (clean rooms),
air heaters or air
coolers, where the
turbulence is very
low.
The probe shall be
installed in the
middle of the duct.
Preferred location
after filters, recti-
fiers, coolers
(no turbulances)
The probe shall be
placed in front of dif-
fusers or confusers.
Filters and coolers
calm down the air
flow.
ALMEMO
®
Manual 3-5-3